Last checked: 13 October 2026.Regulation (EU) 2017/745 on medical devices (MDR)

The rule: Regulation (EU) 2017/745 on medical devices (MDR), Article 27 and Annex VI Part C, and Regulation (EU) 2017/746 on in vitro diagnostics (IVDR), Article 24.

In Switzerland, the Medical Devices Ordinance (MedDO, SR 812.213) applies the same rules and the same dates.

Labelling rule 3 of 5

What the rule asks of a label

Every medical device label carries a Unique Device Identifier, the UDI, in two forms: a barcode a scanner reads, and the same data as text a person reads. The UDI has two parts:

  1. 01UDI-DI, which identifies the manufacturer and the device model. Each packaging level, unit, box and case, has its own.
  2. 02UDI-PI, which identifies the production unit. If the label shows a lot number, serial number, expiry date or software version, that data must be in the UDI-PI.

The barcode follows the rules of one of four issuing entities designated by the EU: GS1, HIBCC, ICCBBA or IFA. With GS1, the UDI is encoded in a GS1 DataMatrix or a GS1-128 code.

Example surgical scissors, curved, 145 mmExample Medical AG, Musterstrasse 1, CH-0000 ExampleUDI(01)00614141000012(11)260115(17)310114(10)A2391(21)SN004212

Example label. The GS1 DataMatrix encodes the data printed next to it.

Four details catch people out:

  • Every packaging level carries its own UDI. Shipping containers are the exception.
  • Where space is tight, the barcode alone is allowed, except on devices used outside healthcare settings, which must keep the readable text.
  • RFID alone is not enough. A device marked with RFID also needs a barcode.
  • The UDI must stay readable for the lifetime of the device, so material and ribbon matter as much as data.
(01) 00614141000012Unit(01)10614141000019Box(01)20614141000016Case

The deadlines

Device classUDI on the label fromDirect marking of reusable devices from
MDR class III and implantable26 May 202126 May 2023
MDR class IIa and IIb26 May 202326 May 2025
MDR class I26 May 202526 May 2027
IVDR class D26 May 2023
IVDR class B and C26 May 2025
IVDR class A26 May 2027

So today every MDR device needs a UDI on its label.

Still to come: IVD class A, and direct marking of reusable class I devices, both on 26 May 2027.

The UDI database EUDAMED has been mandatory for device registration since 28 May 2026. Devices already on the market before that date must be registered by late November 2026. In Switzerland, registration in swissdamed has been mandatory since 1 July 2026, with a transition period to 31 December 2026.

Registration is a database task, not a label task, but the UDI-DI on your label and in the database must match.

Numaco

What our label systems do

  • Build the UDI from your master data, for example from SAP, so the UDI-DI on the label is the same one you registered, and nobody types it.
  • Print the GS1 DataMatrix or GS1-128 code and the matching readable text with the GS1 Application Identifiers: (01) for the UDI-DI, and (10), (11), (17) and (21) for lot, manufacturing date, expiry and serial number.
  • Hold one template per packaging level, each with its own UDI-DI.
  • Match label material and ribbon to the device's life, from autoclave cycles to freezer storage, with the label and ribbon finder.
You

What stays with you

  • Assigning the UDI-DI and Basic UDI-DI, and registering them in EUDAMED and swissdamed.
  • Deciding the device class and so the deadline.
Project responsibilities

Last checked: 13 October 2026.